H3: product analysis: evaluation determined that bur cannot be locked into the attachment.The likely cause of failure is due to distal bearing is broken.It was also noted that the spindle housing is loose, bearings and o-rings are worn, laser markings are faded, color ring is faded and scratched.G3: aware date used as date of analysis performed date as the event is reported based on evaluation findings.This regulatory report is being submitted due to retrospective review through capa 672570.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|