|
The patient was undergoing a thrombectomy procedure in the left m1 segment of the mca using a penumbra system red 72 silver reperfusion catheter (red72 silver), a sendit delivery device (sendit), a benchmark bmx96 access system (bmx96), and a guidewire.The physician placed the bmx96 in the common carotid artery (cca) and performed a contrast run.The red72 silver and the sendit were advanced to the face of the clot.While retracting the sendit from the red72 silver, the physician experienced resistance.The sendit and the red72 silver were removed together.After removal, the physician noticed multiple kinks on the distal end of the red72 silver.The physician advanced a new red72 silver and sendit to the target location.The sendit was removed to initiate aspiration.After three minutes of aspiration no clot was aspirated.The red72 silver was removed and found to be fractured at the distal end.The proximal portion was removed from the patient.The remaining portion extended from the m1 segment to the external carotid artery (eca).Multiple attempts were made to remove the remaining portion of the red72 silver using a snare device.The remaining portion of the red72 silver was left in the patient.There are currently no plans to remove the fractured red72 silver.It was reported that there was improved flow post-procedure.The patient remained in the intensive care unit (icu).On (b)(6) 2026, it was reported that the patient was transferred to a rehabilitation facility and is doing well.
|