Model Number AF03
Medical Device Problem Code Loss of Power (1475)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/29/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H3: product analysis: evaluation of the device could not confirm device stoppage.However, it was noted the internal tube bearings were corroded, fractured and detached, and temperature test could not be performed.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that prior to procedure, the device stops running when it touches bone.There was no patient involved in the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIDAS REX®
Common Device Name
MOTOR, DRILL, PNEUMATIC
Manufacturer (Section D)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
Manufacturer (Section G)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
Manufacturer Contact
justin ellis
8200 coral sea street ne mvs21
mounds view, MN 55112
7635265677
MDR Report Key25700656
Report Number1625507-2026-04142
Device Sequence Number9748098
Product Code HBB
UDI-Device Identifier00721902481864
UDI-Public(01)00721902481864(11)140521
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K020069
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/02/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAF03
Device Catalogue NumberAF03
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/02/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/02/2026
Initial Report FDA Received Date07/03/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/21/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-