The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the cmos module with drive pcb blackout.Based on the result, we concluded that it was caused due to the excessive force applied on the cmos module with drive pcb.In addition, our technician confirmed that the insertion flexible tube fluid damage, the segment fluid damage, the control body fluid damage, the cmos module with drive pcb fluid damage, the light guide cable fluid damage, the lg connector fluid damage, the control body corroded, the lg connector corroded, the operation channel (primary) leak, the operation channel (primary) cut, the remote control buttons cut, and the operation channel (primary) resistance; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report "(b)(4) (image failure)" and/or the risk analysis results, it was evaluated to submit mdr.
|