MEDTRONIC MEXICO S. DE R.L. DE CV SIMUFORM; RING, ANNULOPLASTY Back to Search Results
Model Number 7800RR30
Medical Device Problem Code Intravalvular regurgitation (4068)
Health Effect - Clinical Code Mitral Valve Insufficiency/ Regurgitation (4451)
Date of Event 07/02/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Medtronic received information that during the implant of this 30mm simuform sr annuloplasty heart ring, it was explanted and replaced with an unknown product.The reason for the replacement was reported as mitral regurgitation (mr).After the ring was implanted, a saline test showed poor repair with regurgitation, and the device was therefore removed.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIMUFORM
Common Device Name
RING, ANNULOPLASTY
Manufacturer (Section D)
MEDTRONIC MEXICO S. DE R.L. DE CV
av. paseo del cucapah #10510
tijuana,bc 22210
MX  22210
Manufacturer (Section G)
MEDTRONIC MEXICO S. DE R.L. DE CV
av. paseo del cucapah #10510
tijuana,bc 22210
MX   22210
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key25703818
Report Number9617601-2026-04134
Device Sequence Number13198483
Product Code KRH
UDI-Device Identifier00763000271497
UDI-Public(01)00763000271497(17)290827
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K201449
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/06/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number7800RR30
Device Catalogue Number7800RR30
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/02/2026
Supplement Date Received by Manufacturer07/03/2026
Initial Report FDA Received Date07/03/2026
Supplement Report FDA Received Date07/06/2026
Date Device Manufactured08/27/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening;
Patient Age61 YR
Patient SexMale
Patient Weight71 KG
-
-