BOSTON SCIENTIFIC NEUROMODULATION CORPORATION Infinion? Pro; stimulator, spinal-cord, totally implanted for pain relief Back to Search Results
Model Number SC-2318-50
Medical Device Problem Code Device Fell (4014)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/09/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an implant procedure, the lead was dropped to the floor.There was no harm to the patient.
 
Additional Manufacturer Narrative
D2b: lgw - qrb.
 
Event or Problem Description
It was reported that during an implant procedure, the lead was dropped to the floor.There was no harm to the patient.
 
Additional Manufacturer Narrative
D2b: lgw - qrb.Investigation results, media review, device analysis: the returned lead was analyzed, passed all tests performed, and exhibited normal device characteristics.Device history record: it was confirmed these devices met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Labeling review: review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.Investigation conclusion: based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of unintended use error caused or contributed to event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Infinion? Pro
Common Device Name
stimulator, spinal-cord, totally implanted for pain relief
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION CORPORATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
Boston Scientific Corporation
road 698, lot no. 12
dorado PR 00646 -260
PR   00646-2602
Manufacturer Contact
ricardo amaro castro
25155 rye canyon loop
valencia, CA 91355
+1(320)800-7570
MDR Report Key25707119
Report Number3006630150-2026-04634
Device Sequence Number14732036
Product Code LGW
UDI-Device Identifier00191506018726
UDI-Public(01)00191506018726(17)270919(10)5005630
Combination Product (Y/N)N
Initial Reporter StateNE
Initial Reporter CountryUS
PMA/510(K) Number
P030017
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberSC-2318-50
Device Catalogue NumberSC-2318-50
Device Lot Number5005630
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer06/23/2026
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/09/2026
Supplement Date Received by Manufacturer08/21/2026
Initial Report FDA Received Date07/03/2026
Supplement Report FDA Received Date09/17/2026
Date Device Manufactured09/19/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age79 YR
Patient SexMale
-
-