| Model Number |
SC-2318-50 |
| Medical Device Problem Code |
Device Fell (4014)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/09/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during an implant procedure, the lead was dropped to the floor.There was no harm to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
D2b: lgw - qrb.
|
| |
|
Event or Problem Description
|
|
It was reported that during an implant procedure, the lead was dropped to the floor.There was no harm to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
D2b: lgw - qrb.Investigation results, media review, device analysis: the returned lead was analyzed, passed all tests performed, and exhibited normal device characteristics.Device history record: it was confirmed these devices met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Labeling review: review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.Investigation conclusion: based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of unintended use error caused or contributed to event.
|
| |
|