| Model Number |
9513 |
| Medical Device Problem Codes |
Incorrect, Inadequate or Imprecise Result or Readings (1535); Low Readings (2460)
|
| Health Effect - Clinical Code |
Increased Level of Consciousness (4590)
|
| Date of Event |
05/19/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Patient 5/5: the customer has reported there has been several complaints of the device not recording correctly."also, sedline displaying erroneous psi readings, such as very low psi values despite the patient being awake and able to open their eyes.This has occurred on numerous occasions, and the devices have demonstrated significant inconsistency in their readings.Also, psi values remaining fixed at 25 for a prolonged period despite substantial changes in both surgical stimulation and the level of anaesthesia administered".
|
| |
|
Additional Manufacturer Narrative
|
|
Other, other text: masimo has reached out to the customer to request the return of the device.The device has not been returned to masimo to allow an analysis to be performed.If the device is returned for evaluation or new information is obtained, a follow up report will be submitted.E1.Initial reporter zip code exceeded allowable field length, initial reporter zip code is as follows: m20 4bx.E1.Initial reporter phone number exceeded allowable field length, initial reporter phone number is as follows: (b)(6).
|
| |
|