• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CENTURION VISION SYSTEM; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CENTURION VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065751763
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Wound Dehiscence (1154); Eye Burn (2523); Intraocular Pressure Decreased (4468); Insufficient Information (4580)
Date of Event 06/03/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
 
Event or Problem Description
A nurse reported that during sculpting of nucleus, there was an occlusion of phacoemulsification tip with viscoelastic causing heat to disperse to main incision and the patient experienced corneal burn, collapsing of the anterior chamber, astigmatism and wound leakage and which requires sutures during cataract surgery (with intraocular lens implant).The surgery was completed on the same day with sutures of main wound.The patient symptoms were resolved.This is the one of two patients report received from the same facility.This report pertains to ophthalmic system involved for this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION VISION SYSTEM
Common Device Name
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key25715885
Report Number2028159-2026-00980
Device Sequence Number5141683
Product Code HQC
UDI-Device Identifier00380657517633
UDI-Public(01)00380657517633(17)261006(11)161005(10)NA
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K121555
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 07/06/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number8065751763
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/10/2026
Initial Report FDA Received Date07/06/2026
Date Device Manufactured10/05/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CENTURION SURGICAL PROCEDURE PAK.; HAND PIECE.; UNSPECIFIED OPHTHALMIC VISCOSURGICAL DEVICE.
Outcome Attributed to Adverse Event Other;
Patient Age48 YR
Patient SexMale
-
-