Visual inspection during the investigation, no significant deformations or damage of the valve were determined.Permeability test a permeability test has shown that all components are permeable.Computer controlled test to investigate the claim of under-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valves are tested in both the horizontal as well as the vertical positions.The results show that the valves operate within the permissible tolerance in their respective relevant positions.Adjustment test the progav 2.0 was tested and is adjustable to all specified pressures.Braking force and brake function test the brake functionality test has shown that the brake function operates as expected and the braking force is within the given tolerances.Internal inspection after dismantling of the valves, no deposits were found.Results based on the investigation results, no functional deviations or other abnormalities were detected in either valve.The valves operated within specifications.Visible deposits were detected in the control reservoir.These deposits did not affect the technical specifications at the time of the investigation.Deposits caused by substances naturally present in the body, such as protein, blood, or tissue particles, are among the known and unavoidable risks and side effects of hydrocephalus therapy.Even small amounts of organic deposits can compromise the integrity of the valve.A defect at the time of release can be excluded.The shunt system met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
|