• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. HD AUTOCLAVABLE CAMERA HEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. HD AUTOCLAVABLE CAMERA HEAD Back to Search Results
Model Number CH-S190-XZ-E
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/25/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, the most probable cause of the complaint was traced to damage of switch.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
The customer returned the camera head to the service center for a separate issue.During the device analysis, the service report technician indicates that failed leak test was found.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HD AUTOCLAVABLE CAMERA HEAD
Common Device Name
AUTOCLAVABLE CAMERA HEAD
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key25726560
Report Number3002808148-2026-22884
Device Sequence Number9357926
Product Code FET
UDI-Device Identifier04953170310799
UDI-Public04953170310799
Combination Product (Y/N)N
Initial Reporter CountryHK
PMA/510(K) Number
K102059
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 07/06/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberCH-S190-XZ-E
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/05/2026
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/02/2026
Initial Report FDA Received Date07/06/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/16/2012
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-