• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION 303, INC. Alaris System; Pump, Infusion

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION 303, INC. Alaris System; Pump, Infusion Back to Search Results
Model Number 8015
Medical Device Problem Codes Circuit Failure (1089); Failure to Read Input Signal (1581); Output Problem (3005)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/15/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A follow up report will be submitted with investigation results should the device be repaired or the device/logs be received for evaluation.Per 803.52(f)(11)(iii) the information provided was obtained from servicing activities performed on the device.There were no additional details obtainable or provided at the time of service.
 
Event or Problem Description
It was reported that the device had unit constantly alarms and will not turn off.Had to pull battery and reset pump per tech support.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Alaris System
Common Device Name
Pump, Infusion
Manufacturer (Section D)
CAREFUSION 303, INC.
3750 torrey view ct
san diego CA 92130
Manufacturer (Section G)
CareFusion 303, INC.
10020 pacific mesa blvd
san diego CA 92121 -438
Manufacturer Contact
brett wilko
3750 torrey view ct
san diego, CA 92130
+1(858)617-2000
MDR Report Key25729808
Report Number2016493-2026-43254
Device Sequence Number10645158
Product Code FRN
UDI-Device Identifier10885403519215
UDI-Public(01)10885403519215
Combination Product (Y/N)N
Initial Reporter StateHI
Initial Reporter CountryUS
PMA/510(K) Number
K211218
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/05/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number8015
Device Catalogue Number8015 Alaris PCU 1.5 Module
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer06/30/2026
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/15/2026
Supplement Date Received by Manufacturer08/05/2026
Initial Report FDA Received Date07/07/2026
Supplement Report FDA Received Date08/07/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/20/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-