| Model Number |
H74939171100470 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/15/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D2b - pro code (product code): fge, lit g4 - premarket / 510(k): k141521, k141597 investigation results: investigation conclusion based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of adverse event related to patient condition.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.A 10.0 x 40, 75cm mustang balloon catheter was advanced for dilation.During inflation at 6 atmospheres, the balloon ruptured.The procedure was completed with a different device.There were no patient complications as a result of this event.
|
| |
|