|
A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.The reason for reoperation: capsular contracture baker grade iv.
|
|
Device evaluation: the device related to the reported event capsular contracture was received on september 03, 2026, with lot number 1344019.Per the investigation procedure, the device is analyzed through visual microscopic inspection if openings are observed and a weight verification.Per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required:
¿ capsular contracture: unable to observe through visual inspection as it is a physiological phenomenon.None of the other observations performed during the device analysis creases, wear abrasion and deformation are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.Additional, changed, and/or corrected data: d.9., h.2., h.3., h.6.
|