| Model Number |
6947M55 |
| Medical Device Problem Codes |
Impedance Problem (2950); High Capture Threshold (3266)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/16/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the right ventricular (rv) lead exhibited an increase in pacing impedance, a decrease in impedance on both defi brillation coils, and high capture thresholds.The lead remains in use.It was also reported that the patient received anti-tachycardia pacing (atp) and high voltage therapy from the single chamber implantable cardioverter defibrillator (icd) for possible supra ventricular tachycardia (svt) that the device detected as ventricular fibrillation (vf), with the first shock inducing a true ventricular arrhythmia.The icd remains in use.No further patient complications have been reported as a result of this event.
|
| |
|