| Catalog Number |
318-02 |
| Medical Device Problem Code |
Device Sensing Problem (2917)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/24/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported by critical care nurses in an online survey that a nurse stated the arctic sun temperature management system did not successfully control patient temperature by achieving and maintaining patient to target temperature because "faulty pad, replaced and temperature control was achieved" in relation to "arctic sun 5000" and pads "neonatal".Additionally, in the online survey, in chart 6 a nurse stated that no, the arctic sun temperature management system did not successfully monitor the patient's temperature throughout therapy because "faulty pad" in relation to "arctic sun 5000" and pads "neonatal".
|
| |
|
Event or Problem Description
|
|
It was reported by critical care nurses in an online survey that a nurse stated the arctic sun temperature management system did not successfully control patient temperature by achieving and maintaining patient to target temperature because "faulty pad, replaced and temperature control was achieved" in relation to "arctic sun 5000" and pads "neonatal".Additionally, in the online survey, in chart 6 a nurse stated that no, the arctic sun temperature management system did not successfully monitor the patient's temperature throughout therapy because "faulty pad" in relation to "arctic sun 5000" and pads "neonatal".
|
| |
|
Additional Manufacturer Narrative
|
|
The initial reporter's phone number: (b)(6).The reported event was inconclusive because this investigation did not result in any additional findings and no sample was available for evaluation.The reported event is addressed within the labeling as it states:
attach the pad¿s line connectors to the fluid delivery line manifolds.See arctic sun temperature management system operators manual and help screens for detailed instructions on system use.If the pad fails to prime or a significant continuous air leak is observed in the pad return line, check the connections, then if needed, replace the leaking pad.Once the pad is primed, assure the steady state flow rate displayed on the control panel is appropriate.The minimum flow rate should be 1.0 lmin.A dhr could not be performed since no lot number was provided.No additional actions can be taken at this time.Correction: d.Udi format updated with available product information per gudid.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Search Alerts/Recalls
|