| Model Number |
GIF-1TH190 |
| Medical Device Problem Code |
Device Reprocessing Problem (1091)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: the use of products that contain silicone could cause deposits of white foreign material.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
The customer returned the gastrointestinal videoscope to the service center for evaluation related to a separate issue.During testing, the service technician identified that the device had white foreign objects in the air/water cylinder and tube.There was no patient involvement.
|
| |
|