| Model Number |
429878 |
| Medical Device Problem Code |
Extracardiac stimulation (1463)
|
| Health Effect - Clinical Code |
Inappropriate Device Stimulation of Tissue (4928)
|
| Date of Event |
06/03/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient was post-operatively experiencing diaphragmatic stimulation from the left ventricular (lv) lead.Pacing amplitude and polarity were adjusted with no further reports of stimulation.The lead remains in use.The patient is a participant in a clinical study.No further patient complications have been reported as a result of this event.
|
| |
|