| Catalog Number |
0606560 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/20/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H11: as the lot number for the device was not provided, a review of the device history records could not be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow up report will be submitted as applicable.Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
On (b)(6), 2026, a patient underwent a chronic catheter placement procedure using a hickman cv catheter.During the procedure, a small tear was observed in the blue lumen during flushing, along with a minor trickle of blood from the affected site.There was no reported patient injury.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2026, a patient underwent a chronic catheter placement procedure using a hickman/leonard/broviac central venous catheter.On (b)(6) 2026, following the procedure, a small tear was observed in the blue lumen during flushing, along with a minor trickle of blood from the affected site.There was no reported patient injury.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2026, a patient underwent a chronic catheter placement procedure using a hickman/leonard/broviac central venous catheter.On (b)(6) 2026, following the procedure, a small tear was observed in the blue lumen during flushing, along with a minor trickle of blood from the affected site.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: this supplemental report is submitted to correct an error in the previously submitted reports.The initial mdr listed an incorrect manufacturing location (g1).The correct manufacturing location for this device is bard puerto rico.No new event information is being reported in this supplement.Manufacturing review: a manufacturing review was not requested as the lot number reported is unknown.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.The investigation is inconclusive for the reported tear in blue lumen and fluid leak issues as no objective evidence was provided for review.The definitive root cause could not be determined based upon available information.It is unknown whether patient and/or procedural factors contributed to the event.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.B3, b5, e1, g1, g3, h6 (component, method).Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: this supplemental report is submitted to correct an error in the previously submitted reports.The initial mdr and supplemental report 1 listed an incorrect manufacturing location (g1).The correct manufacturing location for this device is bard shannon limited.No new event information is being reported in this supplement.Manufacturing review: a manufacturing review was not requested as the lot number reported is unknown.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.The investigation is inconclusive for the reported tear in blue lumen and fluid leak issues as no objective evidence was provided for review.The definitive root cause could not be determined based upon available information.It is unknown whether patient and/or procedural factors contributed to the event.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.G1.Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|