| Catalog Number |
51-102050 |
| Medical Device Problem Codes |
Delivered as Unsterile Product (1421); Tear, Rip or Hole in Device Packaging (2385)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).G2: foreign, event occurred in japan.The customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the sterile packaging was found damaged during stock inspection.There was no patient involvement.Attempts have been made and no further information has been provided.
|
| |
|
Event or Problem Description
|
|
No further information is available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated: d4; d9; h2; h3; h4; h6; h10
complaint is not confirmed.Visual examination of the return product identifies that cavities were opened to inspect the precut tear and found neither were torn beyond the precut tear.No other damage was noted.Pouch seal widths were measured, and all the devices met the acceptable 6mm.A device history review is not required.Reported event did not occur in an operating room or as part of medical procedure; medical records are not available for review.Complaint history review was not performed as no problem was found.No problem was found as these are found within specification.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|