STAAR SURGICAL COMPANY EVO/EVO+ VISIAN IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number VICMO_13.2(-15.5)
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 06/08/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H6 - 4581: significant reduction ica, shallowing of ac secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.Claim #(b)(4).
 
Event or Problem Description
The reporter indicated that following an implantable collamer lens (icl) implantation procedure, the patient experienced excessive vault, significant reduction of iridocorneal angles and shallowing of anterior chamber.The lens was exchanged for a same model, power but shorter length lens.The problem was resolved.Cause of the event was reported as unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVO/EVO+ VISIAN IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SUGICAL COMPANY
1911 walker avenue
monrovia 91016
Manufacturer Contact
joselene muniz
1911 walker avenue
monrovia, CA 91016
MDR Report Key25754126
Report Number2023826-2026-01257
Device Sequence Number18678887
Product Code MTA
UDI-Device Identifier00840311302529
UDI-Public(01)00840311302529(11)250724(17)280630
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/08/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Model NumberVICMO_13.2(-15.5)
Device Catalogue NumberVICMO_13.2
Device Lot NumberN/A
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/23/2026
Initial Report FDA Received Date07/08/2026
Date Device Manufactured07/24/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARTRIDGE MODEL-UNK, LOT#-UNK; INJECTOR MODEL-UNK,LOT#-UNK; UNK-MODEL-UNK, LOT#-UNK
Outcome Attributed to Adverse Event Required Intervention;
Patient Age28 YR
Patient SexFemale
-
-