| Model Number |
GIF-1TH190 |
| Medical Device Problem Code |
Failure to Clean Adequately (4048)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/09/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for evaluation.Based on the results of the investigation, olympus confirmed that white foreign material was observed within the device.The material was identified as silicone-based residue/deposits, which may be associated with the use of silicone-containing products such as simethicone or silicone/oil-based lubricants that can remain within the channels even following reprocessing performed in accordance with the instructions for use (ifu).Simethicone and petroleum oil/silicone-based lubricants are non-water soluble and are not recommended for use by olympus.However, a definitive root cause could not be established.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
The customer returned the gastrointestinal videoscope to the service center for evaluation related to a separate issue.During the device analysis, the field service engineer identified that the device exhibited foreign objects within the air/water tube and cylinder.There was no patient involvement.
|
| |
|