ZIMMER BIOMET, INC. G7 OSSEOTI MULTIHOLE 58MM G; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 110010267
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Codes Joint Dislocation (2374); Joint Impingement (4834)
Date of Event 06/18/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).G2: foreign - event occurred in australia.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event or Problem Description
It was reported that the patient underwent an initial hip procedure.During the procedure, zb cup, liner and head was implanted with depuy product.The patient underwent a revision approximately 3 weeks post-implantation due to recurrent dislocations and impinging posterior and inferior on the cup.The cup was revised to be more closed with less anteversion in an effort to account for the impingement and a new freedom construct implanted.Attempts have been made and no further information has been provided.
 
Event or Problem Description
No further information at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were {update/corrected} updated: g3; h2; h3; h6 no product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history records identified no deviations or anomalies during manufacturing.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 OSSEOTI MULTIHOLE 58MM G
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key25763303
Report Number0001825034-2026-02034
Device Sequence Number13196685
Product Code LPH
UDI-Device Identifier00887868356056
UDI-Public(01)00887868356056(17)350704(10)67349331
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K140669
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number110010267
Device Lot Number67349331
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/18/2026
Initial Report FDA Received Date07/09/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/04/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age84 YR
Patient SexFemale
-
-