On (b)(6)2026, a 60-year-old female patient underwent a thrombectomy and atherectomy procedure using a rotarex device in the superficial femoral artery (sfa).During the procedure, it was reported that the rotarex device lost aspiration functionality.The device was withdrawn from the patient and flushed, at which time an abnormality was observed at the catheter tip.The device was removed without difficulty.The procedure was subsequently completed successfully using an alternative device, specifically a laser.There was no reported patient injury.
|
|
H11: manufacturing review: a manufacturing review was conducted and there was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: no physical sample / images / videos were received.A physical investigation was not possible.Due to no sample / images / videos received, the reported malfunction can not be confirmed.A clear root cause could not be identified but a damaged catheter represents a known inherent risk.Labeling review: as the reported event did not allege a labelling or use related issue, a labelling review is not required.H.4.Device manufacture date: unknown.B5, d2b (mcw; dqx), g3, h6 (component, method).Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|