Model Number CYF-V2
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
B3 date of event is unknown.Additional information will be provided if available.The device was returned to olympus for inspection.Based on the results of the investigation, it is likely that the following led to the malfunction: degradation problem; cause traced to component failure.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
The customer returned the cysto-nephro videoscope, which is an olympus asset with no reported complaint.During the evaluation of the device, it was observed the objective lens glue was chipped.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA CYSTO-NEPHRO VIDEOSCOPE
Common Device Name
CYSTO-NEPHRO VIDEOSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key25767398
Report Number3002808148-2026-23386
Device Sequence Number18776795
Product Code FAJ
UDI-Device Identifier04953170339431
UDI-Public04953170339431
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K221683
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 07/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberCYF-V2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/11/2026
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/24/2026
Initial Report FDA Received Date07/09/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/14/2011
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-