| Model Number |
GIF-HQ190 |
| Medical Device Problem Codes |
Electrical /Electronic Property Problem (1198); Melted (1385); Gas/Air Leak (2946); Positioning Problem (3009); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/10/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: cause traced to component failure.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the gastrointestinal videoscope exhibited a burn on the terminal cover.There was no patient involvement.
|
| |
|
Search Alerts/Recalls
|