The product involved in the report has not been returned.A review of the device history record is in-progress.All information reasonably known as of 09 july 2026 has been included in the health authority report.Should additional information be obtained a follow-up health authority report will be provided.The information provided by airlife.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to airlife.Airlife has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the airlife complaint database and is identified as complaint: (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that a airlife product is defective or causes serious injury.
|
|
The product involved in the report has not been returned.A review of the device history record is in-progress.All information reasonably known as of 09 july 2026 has been included in the health authority report.Should additional
information be obtained a follow-up health authority report will be provided.The information provided by airlife.Represents all
of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was
unable or unwilling to provide any further patient, product, or procedural details to airlife.Airlife has no independent knowledge of the event reported but is relaying the information that was provided by the user facility
where the incident occurred.This product incident is documented in the airlife complaint database and is identified as complaint
(b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not
be considered to be an admission that a airlife product is defective or causes serious injury.Device history record (dhr) review:
the device history record for the part number 8150tg-7-50 with lot number 415976 and udi code (b)(4) was reviewed in order to detect any issue related with the reported defect, the complete lot was manufactured, inspected and released on 31mar2023 per our internal procedures and no issues were found.Device history record (dhr) review:
the device history record for the part number 8150tg-7-50 with lot number 514121 and udi code (b)(4) was reviewed in order to detect any issue related with the reported defect, the complete lot was manufactured, inspected and released on 18nov2024 per our internal procedures and no issues were found.Device history record (dhr) review:
the device history record for the part number 8150tg-7-50 with lot number nop031918 and udi code (b)(4) was reviewed in order to detect any issue related with the reported defect, the complete lot was manufactured, inspected and released on 23mar2018 per our internal procedures and no issues were found.Risk assessment
the ultimate risk is low, as that is the highest risk of the patient / caregiver and regulatory / compliance considerations.Complaint does not need to be escalated to carb.Severity: 3/5, serious, ref-20003-c1, f-h30.1.Patient/caregiver risk impact:
there is no applicable risk in a product risk file, so a member of the clinical team was consulted and provided risk id f-h30 with a severity of s = serious (3) and p2 = remote (2) (see attached email).The probability calculated from p1 based on complaints and sales in the previous 24 months (1 complaint(s) / 65,200 units * 100 = 0.0015%) is remote (2), multiplied by p2 from clinical resulting in p = remote (2).By combining p and s, the risk is determined to be low risk (10) per ref-20001-c1 rev 2.2.Regulatory/compliance impact:
disconnection is a minor noncompliance therefore is non-violative (low risk).Low risk - incidents that may likely not lead to legal action by a regulatory authority.Complaint history:
there have been 0 other complaints for the same issue for part number 8150tg-7-50 in the 24 months prior to the reported complaint.P1 based on complaints and sales in the previous 24 months (1 complaint(s) / 65,200 units * 100 = 0.0015%) is remote.
|