| Model Number |
0263 |
| Medical Device Problem Codes |
Defibrillation/Stimulation Problem (1573); Under-Sensing (1661)
|
| Health Effect - Clinical Code |
Syncope/Fainting (4411)
|
| Date of Event |
01/23/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If pertinent information is provided in the future, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient with this implantable cardioverter defibrillator (icd) and right ventricular (rv) lead experienced a syncopal episode.Technical services (ts) was consulted and noted undersensing, which resulted in the rhythm remaining below the rate cutoff (rco) criteria required for therapy delivery.Ts provided troubleshooting recommendations and advised consideration of rv lead revision to optimize sensing performance.This icd and rv lead remain in service.No additional adverse patient effects were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient with this implantable cardioverter defibrillator (icd) and right ventricular (rv) lead experienced a syncopal episode.Technical services (ts) was consulted and noted undersensing, which resulted in the rhythm remaining below the rate cutoff (rco) criteria required for therapy delivery.Ts provided troubleshooting recommendations and advised consideration of rv lead revision to optimize sensing performance.This icd and rv lead remain in service.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.\
\
a review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.There is no indication the device manufacturing process contributed to the reported complaint.\
\
based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of no problem detected.
|
| |
|