• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL Amplatzer Septal Occluder; Transcatheter Septal Occluder

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL Amplatzer Septal Occluder; Transcatheter Septal Occluder Back to Search Results
Model Number 9-ASD-022
Medical Device Problem Code Difficult to Fold, Unfold or Collapse (1254)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/15/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that on (b)(6) 2025 that a 22 mm amplatzer septal occluder was chosen for implant using a 9f amplatzer torqvue delivery system.It was noted that the both the discs of the occluder were observed in cobra shape.The deformed device was never released from the delivery cable while inside the patient.There was no angulation or kink noted in the delivery system.There was no interaction with cardiac structures during deployment.The patient remained hemodynamically stable throughout the procedure.There was no clinically significant delay in procedure and no adverse patient effects reported.
 
Additional Manufacturer Narrative
An event of device deformity could not be confirmed.The device was returned to abbott for investigation and the device met functional specifications when analyzed under non-physiological conditions.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed, and the product met all specifications.Additionally, the lot was reviewed for similar complaints, and there is no indication of a lot-specific product issue.Based on the available information, the cause of the reported device deformity could not be conclusively determined.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Amplatzer Septal Occluder
Common Device Name
Transcatheter Septal Occluder
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
St. Jude Medical #3014918977
edificio #44 calle 0, ave. 2
el coyol alajuela 1897- 4050
CS   1897-4050
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
+1(651)756-5400
MDR Report Key25785474
Report Number2135147-2026-04781
Device Sequence Number13425853
Product Code MLV
UDI-Device Identifier00811806010182
UDI-Public(01)00811806010182(17)300831(10)11045911
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
P000039
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9-ASD-022
Device Catalogue Number9-ASD-022
Device Lot Number11045911
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/01/2026
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/15/2026
Supplement Date Received by Manufacturer09/24/2026
Initial Report FDA Received Date07/10/2026
Supplement Report FDA Received Date09/29/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/16/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-