| Model Number |
9-ASD-030 |
| Medical Device Problem Code |
Difficult to Fold, Unfold or Collapse (1254)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/16/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2025 that a 30 mm amplatzer septal occluder was chosen for implant using a 10f amplatzer torqvue delivery system.It was noted that the both the discs of the occluder were observed in cobra shape.The deformed device was never released from the delivery cable while inside the patient.There was no angulation or kink noted in the delivery system.There was no interaction with cardiac structures during deployment.The patient remained hemodynamically stable throughout the procedure.There was no clinically significant delay in procedure and no adverse patient effects reported.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of device deformity could not be confirmed.The device was returned to abbott for investigation and the device met functional specifications when analyzed under non-physiological conditions.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed, and the product met all specifications.Additionally, the lot was reviewed for similar complaints, and there is no indication of a lot-specific product issue.Based on the available information, the cause of the reported device deformity could not be conclusively determined.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
|
| |
|