The reported event was confirmed cause unknown.The reported failure was able to be reproduced.Visual inspection noted sac burst along lumen.Introduced water into the inflation lumen and did not experience any obstructions to impede flow.Microscopic examination observed no conditions could be associated with the reported event.The labeling is found to be adequate.A review of the lot file showed no rejects or rework associated with this issue.For the affected date code, the rejected units were recorded as in process scrap and will be disposed of after inspection.No, nonconformities or discrepancies were identified in any of the dhrs.Therefore, no additional action required at this time.Corrections: d, g, h.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: sections a through f.The information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|