Model Number NIPG2000
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pain (1994)
Date of Event 06/29/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient experienced shocking sensations during an mri examination.The device was turned off for the scan, and the mri examination was aborted.The patient subsequently turned the device back on and experienced static shock-like sensations throughout the body.The symptoms worsened, and the patient was admitted to the hospital.An hcp assessment regarding the cause of the reported event was not available.The patient later turned the device off again, after which the sensations decreased.No further complications related to this event have been reported.
 
Additional Manufacturer Narrative
A review of the device's manufacturing records was completed, including batch history and device history; there were no deviations or non-conformities associated with the reported event.The device remains implanted, which limits the ability to conduct any physical, functional, and/or root cause analysis.D6a: the exact implant date is unknown.E1.Reporter name: the initial reporter's information is not complete, as the event was reported by a nurse from the hospital.Nevro submits this report in compliance with fda¿s medical device reporting regulations under 21 cfr part 803.Nevro has complied with regulatory investigation requirements and is submitting all information that is reasonably known to us at this time.However, we may not have been able to confirm this information or complete the investigation within the timeframe for filing this report.We may have given no response or an incomplete response to certain questions because we do not currently have information available to provide a complete response.If we later obtain any required information that was not available at the time of this initial report, we will submit a supplemental report.This report is not an admission by anyone that the product described in this report was defective, that it malfunctioned, or that it caused or contributed to the event described in this report.We may conclude that the device had no defect, did not malfunction, or did not cause or contribute to a reportable event.Some of the items on this form include forced-choice terms used by the fda for reporting purposes that do not necessarily reflect nevro¿s conclusions about the causes or nature of the event.This statement should be included with any freedom of information act response.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENZA
Common Device Name
NEVRO SENZA
Manufacturer (Section D)
NEVRO CORP.
1800 bridge parkway
redwood city CA 94065
Manufacturer Contact
john paxton
1800 bridge parkway
redwood city, CA 94065
6504332730
MDR Report Key25787837
Report Number3008514029-2026-00318
Device Sequence Number10656122
Product Code LGW
UDI-Device Identifier00813426020510
UDI-Public00813426020510
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
P130022
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 07/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/11/2020
Device Model NumberNIPG2000
Device Lot Number9443690
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/30/2026
Initial Report FDA Received Date07/10/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/11/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexFemale
-
-