| Medical Device Problem Codes |
Burst Container or Vessel (1074); Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/16/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Block h6: imdrf device code a0402 captures the reportable event of balloon burst.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a cre pro wireguided dase dilatation balloon was being prepared to be used in the esophagus during a balloon dilatation procedure for benign esophageal stenosis performed on (b)(6) 2026.During preparation, the balloon ruptured.No further information has been obtained despite good faith efforts.The procedure was completed with another cre pro wireguided dase dilatation balloon.There were no patient complications reported as a result of this event.The patients condition at the conclusion of the procedure was reported to be stable.Note: the instructions for use (ifu) indicate that this balloon should not be pre-inflated prior to use in the procedure.However, the customer reported that the balloon was pre-inflated outside of the patient.
|
| |
|