• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON LABORATORIES IRELAND LTD. CLAREON PANOPTIX PRO TRIFOCAL TORIC UV ABSORBING IOL W/AUTONOME DELIVERY SYSTEM; LENS, INTRAOCULAR, TORIC OPTICS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON LABORATORIES IRELAND LTD. CLAREON PANOPTIX PRO TRIFOCAL TORIC UV ABSORBING IOL W/AUTONOME DELIVERY SYSTEM; LENS, INTRAOCULAR, TORIC OPTICS Back to Search Results
Model Number PXCAT4
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 06/18/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information provided in d.9., h.3, h.6 and h.11.The product was returned for analysis, and the reported complaint could not be observed.Iol (intraocular lens) was not returned.Only the device was returned.The device was returned in an opened blister tray inside the product carton.The lock-out assembly has been removed.Viscoelastic is observed in the device.The plunger has been nearly fully advanced outside the nozzle.No damage observed.Iol was not returned.The product investigation could not identify the root cause for the reported complaint "the leading haptic was noted to be kinked in the nozzle" as the lens was not returned.Only the device was returned for evaluation.Due to this, we are unable to confirm the lens and the plunger position for advancement and determine a root cause.Based on the results from the manufacturing batch record, the products met release criteria.There have been no other complaints for this batch.Investigation has been completed based on current information.No further action is warranted at this time.Based on our trend analysis, there are no adverse trends identified for this complaint and based on these findings the residual risk remains at an acceptable level.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON PANOPTIX PRO TRIFOCAL TORIC UV ABSORBING IOL W/AUTONOME DELIVERY SYSTEM
Common Device Name
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key25795717
Report Number9612169-2026-01591
Device Sequence Number8981447
Product Code MJP
UDI-Device Identifier00380658491857
UDI-Public(01)00380658491857(17)310425(11)260427(10)26245817
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 07/13/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPXCAT4
Device Catalogue NumberPXCAT4.220
Device Lot Number26245817
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/19/2026
Initial Report FDA Received Date07/13/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/27/2026
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-