LIMACORPORATE S.P.A TEMA ELBOW SYSTEM; HUMERAL STEM H9 LEFT Back to Search Results
Model Number 1504.14.090
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Loose body in joint (4926)
Date of Event 09/14/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Preliminary review of manufacturing records confirmed that all released products were manufactured according to design specifications.The manufacturer will submit a final report as soon as the investigation is completed.
 
Event or Problem Description
Revision surgery was performed on (b)(6) 2021 due to loosening of humeral components of elbow prosthesis.Complete patient clinical history is summarized below.Patient underwent primary surgery in 2006 for fracture when an elbow prosthesis from another manufacturer was implanted.First revision surgery was performed on (b)(6) 2021 with a complete cementless tema assembly consisting of: humeral stem #h7 left (pn 1504.14.070) humeral body large left + screw (pn 1550.15.020) ulnar stem #u6 (pn 1575.14.030 lot 1816847 ster (b)(4) ulnar body large left + screw (pn 1552.15.020 lot 1714267 ster (b)(4) ulnar liner large left (pn 1560.50.020 lot 16at1wl ster (b)(4).Patient bone quality was reported to be poor.After few months, second revision surgery was performed on (b)(6) 2021 for humeral stem loosening (mfr.3008021110-2026-00286) and the above-mentioned humeral components were removed and replaced with: humeral stem #h9 left (pn 1504.14.090 lot 1816474 ster (b)(4) humeral body large left (pn 1550.15.020 lot 1714541 ster (b)(4).No indication regarding the cementation adopted was provided.One month later, on (b)(6) 2021, patient was revised for the third time (hereby reported) for loosening of the recently implanted humeral stem #h9 with a bigger stem (humeral stem #h10 left, pn 1504.14.100) and a new humeral body large left (pn 1550.15.020).Two cerclages around the humerus were implanted and assembly was then linked to pre-existing ulnar component with an axle (axle large pn 1590.15.020).Complete linked assembly lasted almost 2 years, until (b)(6) 2023 when the fourth revision was performed for unknown reasons (mfr 3008021110-2026-00288), and all pre-existing components, except for the humeral stem, were replaced with smaller ones: humeral body small left (pn 1550.15.110) ulnar stem #u5 (pn 1575.14.010) ulnar body small left (pn 1552.14.010) ulnar liner small left (pn 1560.50.010) axle small (pn 1590.15.010).On (b)(6) 2024 the patient underwent a further revision of the humeral assembly (mfr.3008021110-2026-00282), in which the pre-existing humeral stem #h10, implanted 3 years before, was replaced with a smaller one and implant re-linked: humeral stem #h6 left (pn 1504.14.060) humeral body small left (pn 1550.15.110) axle small (pn 1590.15.010).In this case the humeral stem was cemented.In (b)(6) 2026, the treating surgeon requested a custom-made device for patient due to loosened humeral component from infection (complaint (b)(4), mfr.3008021110-2026-00281).Patient has rheumatoid arthritis bone loss.Patient is male, date of birth (b)(6) 1961.The event occurred in the united states.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TEMA ELBOW SYSTEM
Common Device Name
HUMERAL STEM H9 LEFT
Manufacturer (Section D)
LIMACORPORATE S.P.A
via nazionale 52
villanova di san daniele, udine 33038
IT  33038
Manufacturer Contact
via nazionale 52
villanova di san daniele, udine 33038
MDR Report Key25797021
Report Number3008021110-2026-00287
Device Sequence Number6000071
Product Code JDC
UDI-Device Identifier08033390140249
UDI-Public8033390140249
Combination Product (Y/N)N
PMA/510(K) Number
K181362
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/13/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Model Number1504.14.090
Device Lot Number1816474
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date07/13/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age60 YR
Patient SexMale
-
-