JOHNSON & JOHNSON SURGICAL VISION, INC. VERITAS VISION SYSTEM; UNIT, PHACOFRAGMENTATION
|
Back to Search Results |
|
| Model Number |
VRT680300 |
| Medical Device Problem Code |
Suction Problem (2170)
|
| Health Effect - Clinical Code |
Capsular Bag Tear (2639)
|
| Date of Event |
06/15/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during the procedure with the veritas console, the last quadrant did not come to the tip during aspiration.The surgeon flushed the handpiece and tubing, but aspiration still failed to reach the tip.A posterior capsular rupture occurred, necessitating an anterior vitrectomy and insertion of an ar40e intraocular lens.Additional information indicated that that vacuum increased without corresponding aspiration, suggesting a blockage in the handpiece or tubing pack.No further information was provided.Note: this report is against the veritas console.A separate report will be submitted against the handpiece and tubing.
|
| |
|
Additional Manufacturer Narrative
|
|
Section e1: initial reporter: telephone number: (b)(6).Section h3: the device was not returned for evaluation; therefore, a failure analysis of the complaint device cannot be completed.A review of records including device history and complaint trending will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|