| Model Number |
6935M55 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Fracture (1260); Impedance Problem (2950); Unstable Impedance (4085)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/12/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the right ventricular (rv) lead triggered a lead integrity alert (lia) for high rate non-sustained episodes, non physiologic noise on the rv tip to rv ring electrograms (egm), as well as rv tip to rv ring and rv tip to rv coil jumps and variability in impedances.There was a potential fracture noted.The lead remains in use.No patient complications have been reported as a result of this event.
|
| |
|