| Catalog Number |
SRM-375 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/09/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformance's noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "rupture".This record is for the left side.Device was explanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "rupture".This record is for the left side.Device was explanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "rupture".This record is for the left side.Device was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: d9, h3, h6.Device evaluation:
per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required:
- rupture: observed broken device through microscopic inspection assessed as surgical damage and observed a missing piece of shell through microscopic inspection assessed as inconclusive.No further actions are required as it is not related to the manufacturing process.No additional observations performed on the device.No further actions are required.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: h6.
|
| |
|
Search Alerts/Recalls
|