| Catalog Number |
15-421 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/05/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.A review of the device history record has been completed.No deviations or non-conformances noted.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "rupture and capsular contracture baker grade unknown", later healthcare professional reported "capsular contracture baker grade ii".This record is for the left side.Device was explanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "rupture and capsular contracture baker grade unknown", later healthcare professional reported "capsular contracture baker grade ii".This record is for the left side.Device was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: d9, h3, h6.Device evaluation:
per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required:
rupture: observed an opening through microscopic inspection assessed as fold flaw opening.No further actions are required as it is not related to the manufacturing process.Capsular contracture: unable to observe through visual inspection as it is a physiological phenomenon.None of the other observations performed during the device analysis non penetrating nicks are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.
|
| |
|