| Model Number |
72400098 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Mechanical Problem (1384)
|
| Health Effect - Clinical Codes |
Erosion (1750); Urinary Incontinence (4572)
|
| Date of Event |
06/17/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device is not available for analysis; therefore, no physical analysis of the product could be performed.The reported device performance allegation cannot be confirmed.Based on the information available the cause that contributed to the reported event cannot be established.Therefore, a conclusion code of cause not established was assigned to this investigation.Detailed product information was not provided to bsc.We completed a good faith effort to obtain the information.Because the product is unknown at this time, we are unable to provide the complete unique identifier (udi) # and other product specific information.If additional details become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient with this artificial urinary sphincter (aus) experienced recurring urinary incontinence.Upon evaluation of the control pump, no pressure was detected, and a possible minor leak was suspected.Several days later during a follow up hospital visit, the physician observed erosion upon examination of the urethra.Endoscopic images were obtained and provided by the hospital.The device remains implanted and a surgery replacement is planned.No further patient complications were reported.
|
| |
|