• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. Fluency Plus Endovascular Stent Graft

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. Fluency Plus Endovascular Stent Graft Back to Search Results
Catalog Number FEM10080
Medical Device Problem Codes Difficult or Delayed Positioning (1157); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/23/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H11: as the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
On (b)(6) 2026, a patient underwent a stent placement procedure using a fluency device.It was observed that the clear plastic knob on the back of the tuohy was loose and became disconnected, making stent deployment more difficult.There was no reported patient injury.
 
Event or Problem Description
On (b)(6) 2026, a patient underwent a stent placement procedure using a fluency plus endovascular stent graft for the treatment of stenosis at the venous anastomosis via a left upper arm graft.During the procedure, while the stent was being deployed, it was observed that the clear plastic knob on the back of the tuohy was loose and became disconnected, making stent deployment more difficult.The patient is doing well, with no reported complications during or after the procedure.There was no reported patient injury.
 
Additional Manufacturer Narrative
H11: manufacturing review: based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.However, the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the stent delivery system sample was not returned for evaluation and photos were not provided which leads to inconclusive results.In this case, it was reported that a 9fr introducer was used with a 0.035" guidewire for access, the tracking vessel was neither tortuous nor calcified, the lesion was pre-dilated and the device was flushed before use.Based on the provided information and as neither sample nor photos were provided, the investigation is closed with inconclusive results for detachment of the tuohy-borst and difficult to deploy.A definite root cause for the reported event could not be determined.Labeling review: in reviewing the relevant labeling, it was found that the instructions for use (ifu) sufficiently address the potential risks.The ifu state: 'if unusual resistance or high deployment force is encountered during stent graft deployment, abort the procedure, remove the delivery system and use an alternative device.' regarding preparation of the device the ifu state that "prior to loading the vascular system over a guide wire, both ports must be flushed with sterile saline (.).Flushing these lumens will also facilitate stent graft deployment".Regarding accessories the ifu states: "prepare a stiff 0.035" guidewire per its instructions for use and advance the guidewire under fluoroscopy to the target location.The use of an appropriately sized introducer sheath is recommended".The packaging pictograms indicate an introducer size of 9f and a 0.035" guidewire.The ifu states "carefully remove the endovascular system from its packaging and inspect packaging and system for any damage or defects.Do not use if the sterile barrier is compromised".H.4.Device manufacture date: unknown.B5, g3, h6 (result, conclusion).Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Fluency Plus Endovascular Stent Graft
Common Device Name
Endovascular Stent Graft
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
Angiomed GmbH & Co. Medizintechnik KG
wachhausstr. 6
karlsruhe 76227
GM   76227
Manufacturer Contact
michael gaffney
800 w. rio salado pkwy
tempe, AZ 85281
+1(480)303-2689
MDR Report Key25822439
Report Number2020394-2026-01805
Device Sequence Number21435472
Product Code PFV
UDI-Device Identifier04049519008608
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P130029
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberFEM10080
Device Lot NumberANKX3281
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/23/2026
Supplement Date Received by Manufacturer09/17/2026
Initial Report FDA Received Date07/15/2026
Supplement Report FDA Received Date09/22/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-