Venclose, Inc. Venclose System digiRF Generator; Radiofrequency Ablation (RFA) Generator
|
Back to Search Results |
|
| Catalog Number |
VCRFG1 |
| Medical Device Problem Code |
Communication or Transmission Problem (2896)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/23/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the facility had a failed catheter when used with this venclose generator.The generator was confirmed to be in software version 3.35.Additional information was provided stating that during a procedure on (b)(6) 2026, using this venclose generator, a red x error with error 53 was displayed immediately after the catheter was connected.The catheter had not yet been inserted into the vein when the error occurred.The target treatment location was the great saphenous vein (gsv); the affected side was not specified.There was no patient involvement or impact reported.
|
| |
|
Event or Problem Description
|
|
It was reported that the facility had a failed catheter when used with this venclose generator.The generator was confirmed to be in software version 3.35.Additional information was provided stating that during a procedure on (b)(6) 2026, using this venclose generator, a red x error with error 53 was displayed immediately after the catheter was connected.The catheter had not yet been inserted into the vein when the error occurred.The target treatment location was the great saphenous vein (gsv); the affected side was not specified.There was no patient involvement or impact reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: manufacturing review: a dhr and manufacturing review were not required as the event was not classified as an unexpected event, the investigation did not identify any manufacturing related issues, and no repeat failures were found during the service history review.Investigation summary: the venclose generator (b)(6) was received at the bd-bard global service & repair.The venclose generator was visually inspected upon receipt and was found to be in good physical condition.The generator was received with software version 3.35, so it was downgraded to 3.17 for testing.The device was functionally tested and passed the tests during evaluation.The short tests values are 10cm equal to 20.45 v, 2.5cm equal to 6.71 v, 10cm short test equal to pass and 2.5cm short test equal to pass.No other anomalies were identified.Therefore, the investigation is determined to be confirmed for the reported failed to read catheter issue.A definitive root cause for the reported red x software issue on the device could not be determined based upon the provided information.Labeling review: the labeling/packaging review was not required as the reported event is not associated with a labeling, packaging, or use related issue.E1, g3, h6 (method, result, conclusion)
section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: this supplemental report is submitted to correct an error in the previously submitted reports.The initial mdr listed an incorrect manufacturing location (g1).The correct manufacturing location for this device is minnetronix, inc.No new event information is being reported in this supplement.Manufacturing review: a dhr and manufacturing review were not required as the event was not classified as an unexpected event, the investigation did not identify any manufacturing related issues, and no repeat failures were found during the service history review.Investigation summary: the venclose generator serial number (b)(6) was received at the bd-bard global service & repair.The venclose generator was visually inspected upon receipt and was found to be in good physical condition.The generator was received with software version 3.35, so it was downgraded to 3.17 for testing.The device was functionally tested and passed the tests during evaluation.The short tests values are 10cm equal to 20.45 v, 2.5cm equal to 6.71 v, 10cm short test equal to pass and 2.5cm short test equal to pass.No other anomalies were identified.Therefore, the investigation is determined to be confirmed for the reported failed to read catheter issue.A definitive root cause for the reported red x software issue on the device could not be determined based upon the provided information.Labeling review: the labeling/packaging review was not required as the reported event is not associated with a labeling, packaging, or use related issue.G1, g3
section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|