SHIRAKAWA OLYMPUS CO., LTD. CYSTO-NEPHRO VIDEOSCOPE Back to Search Results
Model Number CYF-VHR
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
B3: date of event is unknown.Additional information will be provided if available.The device was returned to olympus for inspection.Based on the results of the investigation, a definitive root cause of the detached device tip could not be determined, however, the issue was likely the result of component failure due to repetitive use stress, handling and or external factors.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
The customer returned the cysto-nephro videoscope to the service center for a separate issue.During the device analysis, the service repair technician found the distal end to be detached, likely due to stress.There was no patient involvement, and no harm was reported as a result of the event.
 
Additional Manufacturer Narrative
This supplemental report is being submitted to provide an update to the following fields: updated fields: d4.
 
Event or Problem Description
No additional information has been provided from the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CYSTO-NEPHRO VIDEOSCOPE
Common Device Name
CYSTO-NEPHRO VIDEOSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key25829829
Report Number3002808148-2026-24131
Device Sequence Number10356235
Product Code FAJ
UDI-Device Identifier04953170310522
UDI-Public04953170310522
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K221683
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 07/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberCYF-VHR
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/07/2026
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/18/2026
Supplement Date Received by Manufacturer07/06/2026
Initial Report FDA Received Date07/15/2026
Supplement Report FDA Received Date07/16/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/04/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-