| Model Number |
9390A-1001 |
| Medical Device Problem Code |
Incorrect Interpretation of Signal (1543)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/24/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Complainant alleged that while attempting to defibrillate a 63-year-old female patient, the device issued a "shock advised" prompt for a heart rhythm clinicians believed was non-shockable.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned to zoll for evaluation; however, the device log was reviewed.The log showed that the device appropriately analyzed the patient's rhythm and advised "no shock advised." no evidence of a device malfunction was identified.The customer collaborated the outcome with tech support over the phone and will not be returning the device since it operated as expected.The customer report is unsubstantiated, and the device operated as intended.Analysis of reports of this type has not identified an increase in trend.
|
| |
|