A review of the device history record has been completed.No deviations or non-conformances noted.The event of "calcification and capsular contracture" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: calcification and capsular contracture, baker grade iii.
|
|
Device evaluation:
per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required:
¿ deflation: observed two openings through microscopic assessed as surgical damage.No further actions are required as it is not related to the manufacturing process.¿ calcification: observed calcification on device through visual inspection.No further actions are required as it is not related to the manufacturing process.None of the other observations performed during the device analysis (crease and wear abrasion) are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.Additional, changed, and/or corrected data: a2, a5, a6, b1, b5, d9, e1, h3, h6.
|