| Model Number |
8621500 |
| Medical Device Problem Codes |
No Device Output (1435); Failure to Deliver (2338)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/23/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the anesthesia machine malfunctioned during operation.The monitor became inoperable (black screen) and ventilation failed.The patient could be ventilated without difficulty using a manual resuscitation bag.No patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation was done on the provided information only.During on-site inspection the main circuit board m48.1 and the microsd card were identified to be faulty and were replaced.In the case of a faulty processor board, the device issues an alarm and stops automatic ventilation.In this state, the procedure can be continued by manually ventilating the patient via the manual ventilation bag.Spontaneous breathing is also possible.After replacement of the m48.1 board and the microsd card the device was tested and confirmed to be operating per manufacturer's specification.The number of similar cases, related to the same root cause, is within the expected range of the respective risk assessment and thus accepted.The field failure rate is subject to permanent monitoring by the responsible product quality board and is rated as acceptable.
|
| |
|