| Catalog Number |
110-420 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/11/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a rupture and exchange from textured to smooth due to concerns with the product.This record is for the right side.The device was explanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a rupture and exchange from textured to smooth due to concerns with the product.This record is for the right side.The device was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: d9, h3, h6.Device evaluation:
per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required:
- rupture: observed two openings through microscopic inspection assessed as unidentified (tear) opening and fold flaw opening.No further actions are required as it is not related to the manufacturing process.None of the other observations performed during the device analysis creases and wear abrasion are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.
|
| |
|