| Model Number |
URF-V2R |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer did not report a malfunction.During inspection of uretero-reno videoscope, missing bending section cover and chipped bending section cover glue was found.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: degradation and stress problems identified.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.Date of event is unknown.Additional information will be provided if available.
|
| |
|
Search Alerts/Recalls
|