• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMERICAN MEDICAL SYSTEMS GreenLight XPS; Powered Laser Surgical Instrument

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMERICAN MEDICAL SYSTEMS GreenLight XPS; Powered Laser Surgical Instrument Back to Search Results
Model Number 0010-0210R
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/01/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that there was a foot pedal malfunction as it keeps stuck on one position.No patient involved.
 
Event or Problem Description
It was reported that there was a foot pedal malfunction as it keeps stuck on one position.No patient involved.
 
Additional Manufacturer Narrative
The device was not returned for analysis; therefore, a technical analysis could not be performed.However, it was serviced onsite by a field service engineer (fse).The reported allegation was confirmed during evaluation.No additional findings were identified.Corrective actions were performed, and the device subsequently underwent functional and safety testing.All test results were satisfactory, and the device was verified to be operating in accordance with manufacturer specifications.The system was confirmed to be fully operational and returned to service.It was confirmed this device met manufacturing specifications prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.A risk review of the greenlight xps was completed and confirmed that the event of foot switch pedal stuck in on position was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established; since the reported malfunction could not be confirmed due to a lack of a returned product and insufficient information to determine the most probable cause for the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GreenLight XPS
Common Device Name
Powered Laser Surgical Instrument
Manufacturer (Section D)
AMERICAN MEDICAL SYSTEMS
innovation ctr, silicon valley
3070 orchard drive
san jose CA 95134 -201
Manufacturer (Section G)
AMERICAN MEDICAL SYSTEMS
3070 orchard drive
san jose CA 95134 -201
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key25849868
Report Number2124215-2026-38375
Device Sequence Number10582152
Product Code GEX
UDI-Device Identifier00878953001906
UDI-Public(01)00878953001906(11)131016(10)XPS50982
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K092735
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0010-0210R
Device Catalogue Number0010-0210R
Device Lot NumberXPS50982
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/03/2026
Supplement Date Received by Manufacturer09/23/2026
Initial Report FDA Received Date07/16/2026
Supplement Report FDA Received Date09/29/2026
Date Device Manufactured10/16/2013
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-