| Model Number |
M302 |
| Medical Device Problem Code |
Nonstandard Device (1420)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/04/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that there were no field allegations made against the performance of this insertable cardiac monitor (icm) device.The device was returned, and device analysis was completed.Patient information unknowns.
|
| |
|
Additional Manufacturer Narrative
|
|
This insertable cardiac monitor (icm) device was thoroughly inspected and analyzed upon receipt at our post market quality assurance laboratory.Neither visual examination nor x-ray examination of the device identified anomalies.Device received with no telemetry.The device case was removed to facilitate inspection and testing of the internal components.The battery was removed and replaced with an external power source.The current drain was measured and found to be normal.Battery lab found no findings against the battery.A review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.There is no indication the device manufacturing process contributed to the reported complaint.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause linked to device but unable to trace more specifically.
|
| |
|