BOSTON SCIENTIFIC CORPORATION LUX-DX II; INSERTABLE CARDIAC MONITOR Back to Search Results
Model Number M302
Medical Device Problem Code Nonstandard Device (1420)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/04/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that there were no field allegations made against the performance of this insertable cardiac monitor (icm) device.The device was returned, and device analysis was completed.Patient information unknowns.
 
Additional Manufacturer Narrative
This insertable cardiac monitor (icm) device was thoroughly inspected and analyzed upon receipt at our post market quality assurance laboratory.Neither visual examination nor x-ray examination of the device identified anomalies.Device received with no telemetry.The device case was removed to facilitate inspection and testing of the internal components.The battery was removed and replaced with an external power source.The current drain was measured and found to be normal.Battery lab found no findings against the battery.A review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.There is no indication the device manufacturing process contributed to the reported complaint.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause linked to device but unable to trace more specifically.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LUX-DX II
Common Device Name
INSERTABLE CARDIAC MONITOR
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key25851187
Report Number2124215-2026-38425
Device Sequence Number5470707
Product Code MXD
UDI-Device Identifier00802526620713
UDI-Public00802526620713
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K231328
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 07/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/23/2026
Device Model NumberM302
Device Catalogue NumberM302
Device Lot Number111824
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/04/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/24/2026
Initial Report FDA Received Date07/17/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/04/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-