Allergan (Costa Rica) Inspira Smooth Silicone Gel Filled Breast Implant; prosthesis, breast, noninflatable, internal, silicone gel-filled Back to Search Results
Catalog Number SSX-800
Medical Device Problem Code Device Appears to Trigger Rejection (1524)
Health Effect - Clinical Codes Capsular Contracture (1761); Hematoma (1884)
Date of Event 06/23/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade iv.
 
Event or Problem Description
Healthcare professional reported \"capsular contracture baker grade iv\", \"patient had an injury\", \"had a bleed\" and \"implant is getting hard\".This record is for left side.The device remains implanted.
 
Additional Manufacturer Narrative
Additional, changed, and/or corrected data: b5, d6b, h6.\ \ the event of hematoma is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.\ \ reason for reoperation: hematoma.
 
Additional Manufacturer Narrative
Additional, changed, and/or corrected data: b5, d9, h3, h6.\ \ device evaluation:\ per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required:\ \ - capsular contracture: unable to observe through visual inspection as it is a physiological phenomenon.\ - hematoma: unable to observe through visual inspection as it is a physiological phenomenon.\ \ none of the other observations performed during the device analysis (creases and deformation) are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.
 
Event or Problem Description
Healthcare professional later reported \"pectorals muscle tear which caused capsular contraction from bleeding.\" device has been explanted.
 
Event or Problem Description
Healthcare professional reported \"capsular contracture baker grade iv\", \"patient had an injury\", \"had a bleed\" and \"implant is getting hard\".Later healthcare professional reported \"pectorals muscle tear which caused capsular contraction from bleeding\".This record is for left side.The device was explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Inspira Smooth Silicone Gel Filled Breast Implant
Common Device Name
prosthesis, breast, noninflatable, internal, silicone gel-filled
Manufacturer (Section D)
Allergan (Costa Rica)
900 parkway global park
zona franca
la aurora de heredia
CS 
Manufacturer (Section G)
Allergan (Costa Rica)
900 parkway global park
zona franca
la aurora de heredia
CS  
Manufacturer Contact
terry ingram
2525 dupont drive
irvine, CA 92612
+1(847)936-6324
MDR Report Key25851549
Report Number9617229-2026-13605
Device Sequence Number5167887
Product Code FTR
UDI-Device Identifier10888628033467
UDI-Public(01)10888628033467(17)Unknown
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/18/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberSSX-800
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/03/2026
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/23/2026
Supplement Date Received by Manufacturer07/23/2026
09/03/2026
Initial Report FDA Received Date07/17/2026
Supplement Report FDA Received Date08/14/2026
09/18/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age44 YR
Patient SexFemale
Patient Weight167 LB
Patient RaceBlack Or African American
-
-